Out Of Stock
Notify me when its in stockLoyka clinical refractometer can measure urea, serum protein, and refractive index. This refractometer model is mostly used in health institutions, veterinary clinics, and laboratories.
Loyka clinical refractometer measures urea in the range of (1000 . . . 1050 sg).
This urea refractometer measures serum protein in the range of (0 . . . 12 gr/100ml).
LOYKA Clinical refractometer measures refractive index in the range of (1333 . . . 1366 nD).
SG (Specific Gravity) means density or "specific weight".
This refractometer model is mostly used in health institutions, veterinary clinics, and laboratories.
This refractometer device requires no power source and is very easy to use.
To read the measurement results more comfortably, make sure the prism receives enough light.
Zero adjustment is done with the calibration liquid that comes with the device, and clarity adjustment is done via the eyepiece.
This refractometer has an ATC feature. ATC compensation range is 10°C to 20°C.
The prism of the refractometer must be protected from shocks and scratches. With a scratch-free prism, you will always get clear results.
For better and clearer results, the measuring prism should be cleaned with a soft cloth after each use.
Accessories included with the refractometer are; calibration liquid, hard carrying case, small screwdriver for calibration, plastic dropper, cleaning cloth, and user manual.
HOW TO MEASURE?
Hold the refractometer by its body so that the inclined prism at the tip, where the sample is dropped, faces the light source.
With the cover closed, place the eyepiece to your eye and try to read the value on the measurement scale you see.
If the values you see are not clear, turn the adjustment ring above the eyepiece to make the measurement scale clearer.
ZERO CALIBRATION:
First, open the prism cover and drop 1-2 drops of distilled water on the inclined prism surface, then close the cover slowly.
This will leave a thin layer of water between the prism surface and the cover.
If the dark-light separation image around the zero line on the measurement scale is not clear;
Adjust the calibration screw with a screwdriver until the dark-light separation aligns exactly with the zero line on the measurement scale.
TECHNICAL SPECIFICATIONS
Measurement Range (Urea) : 1000 . . . 1050 sg
Sensitivity (Urea) : 0.005 sg
Measurement Range (Serum Protein) : 0 . . . 12 gr/100 ml
Sensitivity (Serum Protein) : 0.2 gr/100 ml
Measurement Range (Refractive Index) : 1333 . . . 1366 nD
Sensitivity (Refractive Index) : 0.0005 nD
ATC Feature : Yes
Dimensions : 27 x 40 x 170 mm
Weight : 125 gr.